ISO 6918:1984
A test solution is prepared by one of the two following ways: a) fusion of a test portion with sodium carbonate in the case of total calcium and magnesium contents, and b) from an aliquot portion of "solution A" (ISO 2217) for alkali-soluble calcium and magnesium contents. Lanthanum ions are added to the solution to suppress certain interferences. The solution is aspirated into an acetylene/dinitrogen monoxide flame and the measurement are carried out at a wavelength of 422.7 nm (calcium line) and of 285.2 nm (magnesium line).
SDO:
ISO
Language:
English
ICS Codes:
71.060.50
Status:
Withdrawn
Publish date:
1984-10-31
Standard Number: