CSA C22.2 No. 60601-2-1:23
Scope
Replacement:
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ELECTRON
ACCELERATORS, hereafter referred to as ME EQUIPMENT, in the range 1 MeV to 50 MeV, used for
TREATMENT of PATIENTS.
NOTE 1 While ELECTRON ACCELERATORS used for TREATMENT of PATIENTS are always ME EQUIPMENT, there are times
in this document where they are referred to as EXTERNAL BEAM EQUIPMENT (EBE). Usage of EBE does not remove the
requirements placed on the ME EQUIPMENT but is meant to clarify that the ME EQUIPMENT being discussed is the EBE
and not some other ME EQUIPMENT that may be part of the system configuration.
This particular standard, with the inclusion of TYPE TESTS and SITE TESTS, applies to the
manufacture and some installation aspects of ELECTRON ACCELERATORS and their included
equipment used to increase the precision, accuracy and volumetric targeting of the TREATMENT
delivery
– intended for RADIOTHERAPY in medical practice, including those in which the selection and
DISPLAY of TREATMENT PARAMETERS can be controlled automatically by PROGRAMMABLE
ELECTRONIC SUBSYSTEMS (PESS),
– that, under NORMAL CONDITIONS and in NORMAL USE, deliver a RADIATION BEAM of X-RADIATION
or ELECTRON RADIATION having
• NOMINAL ENERGY in the range 1 MeV to 50 MeV,
• maximum ABSORBED DOSE RATES between 0,001 Gy × s–1 and 1 Gy × s–1 at the ERP from
the RADIATION SOURCE, and
• REFERENCE TREATMENT DISTANCES (RTDs) between 0,5 m and 2 m from the RADIATION
SOURCE;
and
– intended to be
• for NORMAL USE, operated under the authority of the RESPONSIBLE ORGANIZATION by
QUALIFIED PERSONS appropriately licensed or having the required skills for a particular
medical application, for particular SPECIFIED clinical purposes,
• maintained in accordance with the recommendations given in the INSTRUCTIONS FOR USE,
and
• subject to regular QUALITY ASSURANCE performance and calibration checks by a QUALIFIED
PERSON.
NOTE 2 In this particular standard, all references to installation refer to installation in the RESPONSIBLE
ORGANIZATION's premises.
NOTE 3 In this particular standard, all references to ABSORBED DOSE refer to ABSORBED DOSE in water.
NOTE 4 The limits on maximum ABSORBED DOSE RATES are included for two reasons. The first is due to requirements
related to time in this document. This restriction limits the total dose that could be delivered during a SPECIFIED time
(examples: timer, TIME TO INTERRUPT or TERMINATE, LATENCY). The second is to limit the amount of RADIATION damage
that can occur during the time required to take action (often as a follow up to an INTERRUPTION or TERMINATION OF
IRRADIATION). Wherever requirements were made to limit the amount of dose delivered before action is taken, the
RADIATION damage was considered to be independent of the dose rate and only dependent on the dose. This would
largely hold true if the dose rate stayed within the range stated above.
IEC 61217 gives guidance on the designation of ME EQUIPMENT movements, the marking of
scales, their zero positions and the direction of movement with increasing value (see
201.7.4.101).
IEC 60976:2007 specifies methods of testing and disclosure of functional performance of
medical ELECTRON ACCELERATORS and is intended to facilitate comparisons of accelerator-based
ME EQUIPMENTs of different manufacture. IEC 60976:2007 contains no safety requirements, and
is not required to show compliance with this document. Until IEC 60976:2007 and
IEC TR 60977:2008 are updated to match this document, it is suggested that MANUFACTURERS
replace the word "ISOCENTRE" with "EQUIPMENT REFERENCE POINT" when reading the test
methods.
IEC TR 62926 provides guidance for integration of ELECTRON ACCELERATORS with other
equipment.
IEC TR 63183 provides guidance on the construction of error and warning messages.