CSA C22.2 No. 62990-1:20
1 Scope
This Standard applies to equipment that is intended to be installed or used in accordance with
CSA C22.1,Canadian Electrical Code, Part I.
Group I electrical equipment intended for use in mines susceptible to firedamp is regulated by the
provinces and territories of Canada. Such types of equipment (e.g., caplights) should be referred to the
respective authority having jurisdiction.
1 Scope
This part of IEC 62990 specifies general requirements for design, function and performance,
and describes the test methods that apply to portable, transportable, and fixed equipment for
the detection and concentration measurement of toxic gases and vapours in workplace
atmospheres and other industrial and commercial applications.
This document is applicable to continuously sensing equipment whose primary purpose is to
provide an indication, alarm and/or other output function the purpose of which is to indicate
the presence of a toxic gas or vapour in the atmosphere and in some cases to initiate
automatic or manual protective action(s). It is applicable to equipment in which the sensor
generates an electrical signal when gas is present.
This document applies to two types of equipment:
• Type HM (Health Monitoring) ‘occupational exposure’ equipment:
For occupational exposure measurement, the performance requirements are focused on
uncertainty of measurement of gas concentrations in the region of Occupational Exposure
Limit Values (OELV). The upper limit of measurement will be defined by the manufacturer
in accordance with 4.2.1.
• Type SM (Safety Monitoring) ‘general gas detection’ equipment:
For general gas detection applications (e.g. safety warning, leak detection), the
performance requirements are focused on alarm signalling. The upper limit of
measurement will be defined by the manufacturer according to the intended use of the
equipment.
In general, the requirements for accuracy will be higher for Type HM equipment than for Type
SM equipment. The same equipment may meet the requirements of both Type HM and Type
SM.
For equipment used for sensing the presence of multiple gases this document applies only to
the detection of toxic gas or vapour.
This document is not applicable to equipment:
– with samplers and concentrators such as sorbents or paper tape having an irreversible
indication;
– used for the measurement of gases and vapours related to the risk of explosion;
– used for the measurement of oxygen;
– used only in laboratories for analysis or measurement;
– used only for process measurement purposes;
– used in the domestic environment;
– used in environmental air pollution monitoring;
– used for open-path (line of sight) area gas measurement;
– used for ventilation control in car parks or tunnels.