201.1.1 *Scope
Replacement:
This particular standard applies to BASIC SAFETY and ESSENTIAL PERFORMANCE of INVASIVE
BLOOD PRESSURE MONITORING EQUIPMENT as defined in 201.3.63, hereinafter referred to as
ME EQUIPMENT.
This particular standard does not apply to catheter tubing, catheter needles, Luer locks, taps
and tap tables that connect to the DOME.
This particular standard does not apply to NON-INVASIVE BLOOD PRESSURE MONITORING
EQUIPMENT.