201.1 Scope, object and related standards
Clause 1 of the general standard1) applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
ELECTRON ACCELERATORS, hereafter referred to as ME EQUIPMENT, in the range 1 MeV to
50 MeV, used for treatment of PATIENTS.
This particular standard, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively
to the manufacture and some installation aspects of ELECTRON ACCELERATORS
– intended for RADIOTHERAPY in human medical practice, including those in which the
selection and DISPLAY of operating parameters can be controlled automatically by
PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS),
– that, under NORMAL CONDITIONS and in NORMAL USE, deliver a RADIATION BEAM of XRADIATION
and/or ELECTRON RADIATION having
• NOMINAL ENERGY in the range 1 MeV to 50 MeV,
• maximum ABSORBED DOSE RATES between 0,001 Gy × s–1 and 1 Gy × s–1 at 1 m from
the RADIATION SOURCE,
• NORMAL TREATMENT DISTANCES (NTDs) between 0,5 m and 2 m from the RADIATION
SOURCE,
and
– intended to be
• for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED
PERSONS by OPERATORS having the required skills for a particular medical application,
for particular specified clinical purposes, e.g. STATIONARY RADIOTHERAPY or MOVING
BEAM RADIOTHERAPY,
• maintained in accordance with the recommendations given in the INSTRUCTIONS FOR
USE,
• subject to regular quality assurance performance and calibration checks by a
QUALIFIED PERSON.
NOTE 1 In this particular standard, all references to installation refer to installation in the RESPONSIBLE
ORGANIZATION’S premises.
NOTE 2 In this particular standard, all references to ABSORBED DOSE refer to ABSORBED DOSE in water.
IEC 61271 gives guidance on the designation of ME EQUIPMENT movements; the marking of
scales, their zero positions and the direction of movement with increasing value (see
201.7.4.101).
IEC 60676 specifies methods of testing and disclosure of functional performance of medical
ELECTRON ACCELERATORS. The standard is intended to facilitate comparisons of acceleratorbased
ME EQUIPMENTs of different manufacture. IEC 60676 contains no safety requirements,
and is therefore not required for compliance with this particular standard. It should also be
noted (as stated in the Introduction to IEC 60976:2007) that tests specified in IEC 60976 are
not necessarily appropriate for ensuring that any individual medical ELECTRON ACCELERATOR
conforms to the declared functional performance during the course of its working lifetime.
NOTE 3 IEC/TR 60977, Medical electrical equipment – Medical electron accelerators – Guidelines for functional
performance characteristics, is a related technical report that provides performance guidelines. It shall not be
construed as a standard.