Safety requirements for electrical equipment for measurement, control, and laboratory use — Part 2-091: Particular requirements for cabinet X-ray systems

Logo
CSA Group
Standards Development Organisation:
Working Program:
Designation Number:
CSA C22.2 NO. 61010-2-091:19
Standard Type:
National Standard of Canada - Adoption of International Standard
Standard Development Activity:
Reaffirmation
ICS code(s):
19.080
71.040.10
Status:
Open for SDO comment
SDO Comment Period Start Date:
SDO Comment Period End Date:
Posted On:

Scope:

Scope

This clause of Part 1 is applicable, except as follows:

 

1.1 Scope

 

1.1.1 Equipment included in scope

 

Deletion:

 

Delete the first paragraph.

 

Replacement:

 

Replace the second paragraph (above items a) to c)) with the following new text:

 

This part of IEC 61010 specifies particular safety requirements for cabinet X-ray systems, which fall under any of categories a), b) or c) below.

 

Addition:

 

Add the two following new paragraphs at the end of the subclause:

 

Equipment covered by this document can be both PROTECTED EQUIPMENT or PARTIALLY PROTECTED EQUIPMENT, with X-ray generator voltage up to 500 kV.

 

A cabinet X-ray system is a system that contains an X-ray tube installed in a cabinet, which, independently of existing architectural structures except the floor on which it may be placed, is intended to contain at least that portion of a material being irradiated, provide radiation attenuation and prevent operator access to the radiation beam, during generation of X-radiation.

 

These cabinet X-ray systems are used in industrial, commercial, and public environments, for example, to inspect materials, to analyse materials, and to screen baggage.

 

1.1.2 Equipment excluded from scope

 

Addition:

 

Add the following new items to the list:

 

aa) Equipment intended to apply X-radiation to humans or animals;

 

bb) Equipment incorporating an X-ray tube but not incorporating complete shielding against X-radiation HAZARDS, such as:

 

– equipment intended to be used within a shielded room which excludes personnel during operation;

 

– equipment intended to be used with separate portable or temporary shielding;

 

– equipment intended to produce an emerging beam of X-radiation.

 

1.2 Object

 

1.2.1 Aspects included in scope

 

Addition:

 

Add the following new text to the end of the first paragraph:

 

This part of IEC 61010 specifies requirements for the design and methods of construction of cabinet X-ray systems to provide adequate protection for OPERATORS, bystanders, trained service personnel and the surrounding area against unintentionally-emitted X-radiation and from mechanical HAZARDS related to their conveyors.

Project need:

Project Need

To review the Standard within the required 5 year period.

Note: The information provided above was obtained by the Standards Council of Canada (SCC) and is provided as part of a centralized, transparent notification system for new standards development. The system allows SCC-accredited Standards Development Organizations (SDOs), and members of the public, to be informed of new work in Canadian standards development, and allows SCC-accredited SDOs to identify and resolve potential duplication of standards and effort.

Individual SDOs are responsible for the content and accuracy of the information presented here. The text is presented in the language in which it was provided to SCC.