Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-061: Particular requirements for laboratory atomic spectrometers with thermal atomization and ionization
Scope:
This clause of Part 1 is applicable except as follows:
1.1 Scope
1.1.1 Equipment included in scope
Replacement:
Replace the text, except the first paragraph, with the following new text:
This part of IEC 61010 applies to electrically powered laboratory atomic spectrometers with thermal atomization.
NOTE 1 Examples include atomic absorption spectrometers, emission flame photometers, atomic fluorescence spectrophotometers, inductively coupled plasma spectrometers, microwave coupled plasma spectrometers and mass spectrometers, all with thermal atomization and ionization (including tubing and connectors which are provided by the manufacturer for connection to external supplies).
NOTE 2 If all or part of the equipment falls within the scope of one or more other Part 2 documents of IEC 61010 as well as within the scope of this document, consideration is given to those other Part 2 documents.
1.1.2 Equipment excluded from scope
Addition:
Add, before the first paragraph, the following new text:
This document does not apply to thermal atomization detectors (flame ionization detectors) used in gas chromatography.
Project need:
To review the Standard within the required 5 year period.
Note: The information provided above was obtained by the Standards Council of Canada (SCC) and is provided as part of a centralized, transparent notification system for new standards development. The system allows SCC-accredited Standards Development Organizations (SDOs), and members of the public, to be informed of new work in Canadian standards development, and allows SCC-accredited SDOs to identify and resolve potential duplication of standards and effort.