Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating of materials

Designation Number:
CAN/CSA-C22.2 NO. 61010-2-010:15
Standard Type:
National Standard of Canada - Adoption of International Standard
Standard Development Activity:
Reaffirmation
Status:
Proceeding to development
SDO Comment Period Start Date:
SDO Comment Period End Date:
Posted On:

Scope:

Scope

This clause of Part 1 is applicable except as follows:

 

1.1.1

 

Equipment included in scope

 

Replacement:

 

This part of IEC 61010 specifies safety requirements for electrically powered laboratory equipment for the heating of materials, where the heating of materials is one of the functions of the equipment.

 

Note - If all or part of the equipment falls within the scope of one or more other part 2 standards of IEC 61010 as well as within the scope of this standard, it will also need to meet the requirements of those other part 2 standards. In particular, if equipmen is intended to be used for IVD purposes, it will need to meet the requirements of IEC 61010-2-101.

 

1.1.2

 

Equipment excluded from scope

 

Addition after item j):

 

a) equipment for the heating and ventilation of laboratories;

b) sterilizing equipment;

c) heating and/or cooling equipment which the OPERATOR is intended to enter, and which is large enough for the OPERATOR to remain inside with the door or doors closed

Project need:

Project Need
To review the Standard within the required 5 year period

Note: The information provided above was obtained by the Standards Council of Canada (SCC) and is provided as part of a centralized, transparent notification system for new standards development. The system allows SCC-accredited Standards Development Organizations (SDOs), and members of the public, to be informed of new work in Canadian standards development, and allows SCC-accredited SDOs to identify and resolve potential duplication of standards and effort.