PROCESSING OF HEALTH CARE PRODUCTS — INFORMATION TO BE PROVIDED BY THE MEDICAL DEVICE MANUFACTURER FOR THE PROCESSING OF MEDICAL DEVICES — PART 2: NON-CRITICAL MEDICAL DEVICES
Scope:
This document specifies requirements for the information to be provided by the medical device manufacturer for the processing of non-critical medical devices not intended to be sterilized (i.e. a medical device that is intended to come into contact with intact skin only or a medical device not intended for direct patient contact).
This includes information for processing prior to use or reuse of the medical device.
Processing instructions are not defined in this document. Rather, this document specifies requirements to assist manufacturers of medical devices in providing detailed processing instructions that consist of the following activities, where applicable:
a) preparation before processing;
b) cleaning;
c) disinfection;
d) drying;
e) inspection and maintenance;
f) packaging;
g) storage;
h) transportation.
This document excludes processing of:
1) critical and semi-critical medical devices;
2) medical devices intended to be sterilized;
3) textile medical devices used in patient draping systems or surgical clothing;
4) medical devices specified by the manufacturer for single use only and supplied ready for use.
Project need:
Note: The information provided above was obtained by the Standards Council of Canada (SCC) and is provided as part of a centralized, transparent notification system for new standards development. The system allows SCC-accredited Standards Development Organizations (SDOs), and members of the public, to be informed of new work in Canadian standards development, and allows SCC-accredited SDOs to identify and resolve potential duplication of standards and effort.