Medical electrical equipment – Part 2-35: Particular requirements for the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use

Designation Number:
C22.2 NO. 60601-2-35
Standard Type:
National Standard of Canada - Adoption of International Standard
Standard Development Activity:
New Standard
ICS code(s):
11.140
11.140.01
Status:
Proceeding to development
SDO Comment Period Start Date:
SDO Comment Period End Date:
Posted On:

Scope:

Scope

201.1 Scope, object and related standards

Clause 1 of the general standard1 applies, except as follows:

201.1.1 * Scope
Replacement:

This part of 60601 International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HEATING DEVICES using BLANKETS, PADS or MATTRESSES in medical use, also referred to as ME EQUIPMENT. HEATING DEVICES intended to prewarm a bed are included in the scope of this document.

If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.

If a clause or subclause is specifically intended to be applicable to a specifically defined type of ME EQUIPMENT, as is the case with FORCED AIR DEVICES, then the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.

HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document, except in 7.2.13 and 8.4.1 of the general standard.

NOTE See also 4.2 of the general standard.

This document does not apply to:

– HEATING DEVICES intended for physiotherapy;
– INFANT RADIANT WARMERS; for information, see IEC 60601-2-21 [1]2;
– INFANT INCUBATORS; for information, see IEC 60601-2-19 [2];
– INFANT TRANSPORT INCUBATORS, for information, see IEC 60601-2-20 [3];
– cooling devices.

Project need:

Project Need
To maintain alignment with international requirements on the particular requirements for the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use. This proposed New Standard is being developed at the request of medical equipment suppliers, certification bodies, and the Canadian Health Industries. It will provide the industry with the latest international and national requirements on the basic safety and essential performance of heating devices using blankets, pads or mattresses and intended for heating in medical use. This new standard will also replace the existing CSA C22.2 No. 80601-2-35 Standard with the same title.

Note: The information provided above was obtained by the Standards Council of Canada (SCC) and is provided as part of a centralized, transparent notification system for new standards development. The system allows SCC-accredited Standards Development Organizations (SDOs), and members of the public, to be informed of new work in Canadian standards development, and allows SCC-accredited SDOs to identify and resolve potential duplication of standards and effort.

Individual SDOs are responsible for the content and accuracy of the information presented here. The text is presented in the language in which it was provided to SCC.